In shoulder arthroplasty, mismatch is defined as the difference in the radius or diameter of curvature between the humeral head and glenoid components. Recommendations for mismatch have not been substantiated scientifically. The purpose of this study was to evaluate the effect of mismatch on glenoid radiolucent lines.
A significant linear relationship was found between mismatch and the glenoid radiolucency score (p < 0.0001), with significantly lower (better) radiolucency scores associated with radial mismatches of >5.5 mm.
In this study of glenohumeral prosthetic mismatch ranging from 0 to 10 mm, the mismatch had a significant influence on the scores for the glenoid radiolucent lines, which were best when the radial mismatch was between 6 and 10 mm. The theoretical risk of prosthetic instability with larger mismatch values was not demonstrated within the range of mismatch values evaluated in this series.
Copyright remains with the authors and publisher.
In shoulder arthroplasty, mismatch is defined as the difference in the radius or diameter of curvature between the humeral head and glenoid components. Recommendations for mismatch have not been substantiated scientifically. The purpose of this study was to evaluate the effect of mismatch on glenoid radiolucent lines.
A significant linear relationship was found between mismatch and the glenoid radiolucency score (p < 0.0001), with significantly lower (better) radiolucency scores associated with radial mismatches of >5.5 mm.
In this study of glenohumeral prosthetic mismatch ranging from 0 to 10 mm, the mismatch had a significant influence on the scores for the glenoid radiolucent lines, which were best when the radial mismatch was between 6 and 10 mm. The theoretical risk of prosthetic instability with larger mismatch values was not demonstrated within the range of mismatch values evaluated in this series.