Component alignment can influence implant longevity as well as perhaps pain and function after unicompartmental knee arthroplasty (UKA), but correct alignment is not consistently achieved. To increase the likelihood that good alignment will be achieved during surgery, smart tools such as robotics or patient-specific instrumentation (PSI) have been introduced.
We hypothesized that UKA performed with PSI would result in improved level gait as ascertained with three-dimensional analysis, implant positioning, and patient-reported outcomes measured by a validated scoring system when compared with conventional instrumentation 3 months and 1 year after surgery.
Our observations suggest that PSI may confer small, if any, advantage in alignment, pain, or function after UKA. This argument can therefore not be used to justify the extra cost and uncertainty related to this technique.
Level of evidence: Level I, therapeutic study. Copyright remains with the authors and publisher.
Component alignment can influence implant longevity as well as perhaps pain and function after unicompartmental knee arthroplasty (UKA), but correct alignment is not consistently achieved. To increase the likelihood that good alignment will be achieved during surgery, smart tools such as robotics or patient-specific instrumentation (PSI) have been introduced.
We hypothesized that UKA performed with PSI would result in improved level gait as ascertained with three-dimensional analysis, implant positioning, and patient-reported outcomes measured by a validated scoring system when compared with conventional instrumentation 3 months and 1 year after surgery.
Our observations suggest that PSI may confer small, if any, advantage in alignment, pain, or function after UKA. This argument can therefore not be used to justify the extra cost and uncertainty related to this technique.
Level of evidence: Level I, therapeutic study. Copyright remains with the authors and publisher.