During shoulder replacement surgery, the normal height of the proximal part of the humerus relative to the tuberosities frequently is not restored because of differences in prosthetic geometry or problems with surgical technique. The purpose of the present study was to determine the effect of humeral prosthesis height on range of motion and on the moment arms of the rotator cuff muscles during glenohumeral abduction.
If a humeral head prosthesis is placed too high relative to the tuberosities, shoulder function is impaired by two potential mechanisms: (1) the inferior capsule becomes tight at lower abduction angles and limits abduction, and (2) the center of rotation is displaced upward in relation to the line of action of the rotator cuff muscles, resulting in smaller moment arms and decreased abduction moments of the respective muscles.
Clinical Relevance: In patients managed with shoulder replacement surgery, limitation of range of motion, loss of abduction strength, and overload with long-term failure of the supraspinatus tendon are potential consequences of positioning the humeral head of the prosthesis proximal to the anatomic position.
Copyright remains with the authors and publisher.
During shoulder replacement surgery, the normal height of the proximal part of the humerus relative to the tuberosities frequently is not restored because of differences in prosthetic geometry or problems with surgical technique. The purpose of the present study was to determine the effect of humeral prosthesis height on range of motion and on the moment arms of the rotator cuff muscles during glenohumeral abduction.
If a humeral head prosthesis is placed too high relative to the tuberosities, shoulder function is impaired by two potential mechanisms: (1) the inferior capsule becomes tight at lower abduction angles and limits abduction, and (2) the center of rotation is displaced upward in relation to the line of action of the rotator cuff muscles, resulting in smaller moment arms and decreased abduction moments of the respective muscles.
Clinical Relevance: In patients managed with shoulder replacement surgery, limitation of range of motion, loss of abduction strength, and overload with long-term failure of the supraspinatus tendon are potential consequences of positioning the humeral head of the prosthesis proximal to the anatomic position.
Copyright remains with the authors and publisher.