Aseptic tibial loosening following primary total knee arthroplasty persists despite technique and device-related advancements. The mechanisms for this mode of failure are not well understood. We hypothesized that knee movement while the cement was curing dispersed lipids at the implant-cement interface and would result in decreased tibial fixation strength.
Knee motion during cement polymerization is associated with significant decreases in tibial implant fixational strength. Reduction in implant pull-out strength was identified with each implant design with motion and varied between designs. Across all tested designs, we recommend limiting motion while cementing the tibial implant to improve fixation strength.
Copyright remains with the authors and publisher.
Aseptic tibial loosening following primary total knee arthroplasty persists despite technique and device-related advancements. The mechanisms for this mode of failure are not well understood. We hypothesized that knee movement while the cement was curing dispersed lipids at the implant-cement interface and would result in decreased tibial fixation strength.
Knee motion during cement polymerization is associated with significant decreases in tibial implant fixational strength. Reduction in implant pull-out strength was identified with each implant design with motion and varied between designs. Across all tested designs, we recommend limiting motion while cementing the tibial implant to improve fixation strength.