Modular dual-mobility (MDM) total hip arthroplasty (THA) implants have an additional metal-metal interface between the metal liner and outer metal shell that poses a risk of corrosion. The purpose of this study is to evaluate retrieved MDM liners to evaluate qualitative and quantitative damage and corrosion patterns at this interface.
Visual and dimensional analysis of all 12 retrieved MDM implants demonstrated identifiable corrosion/wear of the cobalt-chromium metal liner taper of varying severity. These implants should be used judiciously until larger series with clinical correlation can be completed.
Copyright remains with the authors and publisher.
Modular dual-mobility (MDM) total hip arthroplasty (THA) implants have an additional metal-metal interface between the metal liner and outer metal shell that poses a risk of corrosion. The purpose of this study is to evaluate retrieved MDM liners to evaluate qualitative and quantitative damage and corrosion patterns at this interface.