Higher revision rates have been reported in patients who have undergone unicompartmental knee arthroplasty compared with patients who have undergone total knee arthroplasty, with poor component positioning identified as a factor in implant failure. A robotic-assisted surgical procedure has been proposed as a method of improving the accuracy of component implantation in arthroplasty. The aim of this prospective, randomized, single-blinded, controlled trial was to evaluate the accuracy of component positioning in unicompartmental knee arthroplasty comparing robotic-assisted and conventional implantation techniques.
139 patients were randomly assigned to treatment with either a robotic-assisted surgical procedure using the MAKO Robotic Interactive Orthopaedic Arm (RIO) system or a conventional surgical procedure using the Oxford Phase-3 unicompartmental knee replacement with traditional instrumentation. A postoperative computed tomographic scan was performed at 3 months to assess the accuracy of the axial, coronal, and sagittal component positioning.
Robotic-assisted surgical procedures with the use of the MAKO RIO lead to improved accuracy of implant positioning compared with conventional unicompartmental knee arthroplasty surgical techniques.
Level of evidence: Therapeutic Level I . Copyright remains with the authors and publisher.
Higher revision rates have been reported in patients who have undergone unicompartmental knee arthroplasty compared with patients who have undergone total knee arthroplasty, with poor component positioning identified as a factor in implant failure. A robotic-assisted surgical procedure has been proposed as a method of improving the accuracy of component implantation in arthroplasty. The aim of this prospective, randomized, single-blinded, controlled trial was to evaluate the accuracy of component positioning in unicompartmental knee arthroplasty comparing robotic-assisted and conventional implantation techniques.
139 patients were randomly assigned to treatment with either a robotic-assisted surgical procedure using the MAKO Robotic Interactive Orthopaedic Arm (RIO) system or a conventional surgical procedure using the Oxford Phase-3 unicompartmental knee replacement with traditional instrumentation. A postoperative computed tomographic scan was performed at 3 months to assess the accuracy of the axial, coronal, and sagittal component positioning.
Robotic-assisted surgical procedures with the use of the MAKO RIO lead to improved accuracy of implant positioning compared with conventional unicompartmental knee arthroplasty surgical techniques.