The purpose of this study was to determine the efficacy of proximal humerus locking plates (PHLP) and to clarify predictors of loss of fixation.
Design: Retrospective review of patients with proximal humerus fractures fixed with a PHLP.
Setting: Five Level 1 trauma centers.
Patients: 153 patients (111 female, 42 male) 18 years or older with a displaced fracture or fracture-dislocation of the proximal humerus treated with a PHLP between January 1, 2001 and July 31, 2005.
Intervention:
Main Outcome Measures: Statistical analysis was done to establish correlations between loss of fixation and postoperative head-shaft angle in the true AP radiograph, patient age, fracture type, trauma mechanism, number of locking head screws, and type of plate.
This series presents the experience using PHLP in 5 Level 1 trauma centers. There were no intraoperative complications related to the locking plate systems. Despite the use of fixed-angle devices, loss of fixation occurred, primarily in the presence of varus malreduction. Our findings suggest that avoiding varus should substantially decrease the risk of postoperative failures.
Copyright remains with the authors and publisher.
The purpose of this study was to determine the efficacy of proximal humerus locking plates (PHLP) and to clarify predictors of loss of fixation.
Design: Retrospective review of patients with proximal humerus fractures fixed with a PHLP.
Setting: Five Level 1 trauma centers.
Patients: 153 patients (111 female, 42 male) 18 years or older with a displaced fracture or fracture-dislocation of the proximal humerus treated with a PHLP between January 1, 2001 and July 31, 2005.
Intervention:
Main Outcome Measures: Statistical analysis was done to establish correlations between loss of fixation and postoperative head-shaft angle in the true AP radiograph, patient age, fracture type, trauma mechanism, number of locking head screws, and type of plate.
This series presents the experience using PHLP in 5 Level 1 trauma centers. There were no intraoperative complications related to the locking plate systems. Despite the use of fixed-angle devices, loss of fixation occurred, primarily in the presence of varus malreduction. Our findings suggest that avoiding varus should substantially decrease the risk of postoperative failures.