To compare the surgical complications and functional outcome of the Gamma nail intramedullary fixation device versus the Richards sliding hip screw and plate device in intertrochanteric femoral fractures.
Design: A prospective, randomised controlled clinical trial with observer blinding.
Setting: A regional teaching hospital in the United Kingdom.
Patients: All patients admitted from the local population with intertrochanteric fractured femurs were included. There were 400 patients entered into the study and 399 followed-up to one year or death.
Intervention: The devices were assigned by randomization to either a short-type Gamma nail (203 patients) or a Richard's-type sliding hip screw and plate (197 patients).
Main Outcome Measures: The main surgical outcome measurements were fixation failure and reoperation. A functional outcome of pain, mobility status, and range of movement were assessed until one year.
The use of an intramedullary device in the treatment of intertrochanteric femoral fractures is still associated with a higher but nonsignificant risk of postoperative complications. Routine use of the Gamma nail in this type of fracture cannot be recommended over the current standard treatment of dynamic hip screw and plate.
Copyright remains with the authors and publisher.
To compare the surgical complications and functional outcome of the Gamma nail intramedullary fixation device versus the Richards sliding hip screw and plate device in intertrochanteric femoral fractures.
Design: A prospective, randomised controlled clinical trial with observer blinding.
Setting: A regional teaching hospital in the United Kingdom.
Patients: All patients admitted from the local population with intertrochanteric fractured femurs were included. There were 400 patients entered into the study and 399 followed-up to one year or death.
Intervention: The devices were assigned by randomization to either a short-type Gamma nail (203 patients) or a Richard's-type sliding hip screw and plate (197 patients).
Main Outcome Measures: The main surgical outcome measurements were fixation failure and reoperation. A functional outcome of pain, mobility status, and range of movement were assessed until one year.
The use of an intramedullary device in the treatment of intertrochanteric femoral fractures is still associated with a higher but nonsignificant risk of postoperative complications. Routine use of the Gamma nail in this type of fracture cannot be recommended over the current standard treatment of dynamic hip screw and plate.
Copyright remains with the authors and publisher.